24. Other Input Related Spurious Responses, HP 8591EM
16.For each of the frequencies listed in Table
a.Set the synthesized sweeper to the listed CW frequency by pressing
4RECALL5 1 for a CW frequency of 542.8 MHz or 4RECALL5 2 for a CW frequency of 1142.8 MHz.
b.Press 4SGL SWP5 and wait for the completion of a new sweep.
c.On the EMC analyzer, press 4MKR !5 MKR ! HIGH and record the
The Actual MKR 1 Amplitude should be greater than the Minimum MKR 1 Amplitude listed in the table below.
Note that the Minimum MKR 1 Amplitude is 10 dB more positive than the specication. This is due to the 10 dB change in reference level made in step 15.
Table
Synthesized Sweeper | TR |
| Actual MKR 1 | Minimum MKR 1 | |
CW Frequency | Entry |
| Amplitude (dBc) | Amplitude (dBc) | |
|
|
|
|
|
|
542.8 MHz | 1 |
|
|
| 55 |
1142.8 MHz | 2 |
|
|
| 55 |
|
|
| |||
|
|
|
|
|
|
17.Record both of the Actual MKR 1 Amplitude readings listed in Table