5. Frequency Readout and Marker Count Accuracy, HP 8593EM, HP 8595EM, and HP 8596EM

5.Frequency Readout and Marker Count A ccuracy, HP 8593EM, HP 8595EM, and HP 8596EM

This performance veri￿cation test must be performed with the EMC analyzer set in the spectrum analyzer mode.

The frequency readout accuracy of the EMC analyzer is tested with an input signal of known frequency. By using the same frequency standard for the EMC analyzer and the synthesized sweeper, the frequency reference error is eliminated.

Equipment Required

Synthesized sweeper

Adapter, Type N (f) to APC 3.5 (m)

Adapter, APC 3.5 (f) to APC 3.5 (f)

Cable, APC 3.5, 91 cm (36 in)

Cable, BNC, 122 cm (48 in)

Additional Equipment for Option 026

Adapter, 3.5 mm (f) to 3.5 mm (f)

Figure 2-5.

Frequency Readout Accuracy Test Setup,

HP 8593EM, HP 8595EM, and HP 8596EM

Performance Veri￿cation Tests 2-17