9. Frequency Span Readout Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
More 1 of 3 More 2 of 3 MARKER 1
More 3 of 3 NEXT PEAK
16.Record the
17.Repeat steps 15 and 16 for the 300 Hz EMC analyzer span setting.
18.Verify that the 300 Hz span setting is within 225 Hz to 255 Hz.
Table
EMC Analyzer | Synthesizer/Level | Synthesized Sweeper |
| |||
Span Setting | Generator | Frequency |
|
|
|
|
| Frequency |
|
|
|
|
|
|
|
|
|
|
| |
| MHz | MHz | Min. | TR Entry | Max. | |
|
|
|
|
|
|
|
10.10 MHz | 66.000 | 74.000 | 7.70 | MHz | 2 | 8.30 MHz |
10.00 MHz | 66.000 | 74.000 | 7.80 | MHz | 3 | 8.20 MHz |
100.00 kHz | 69.960 | 70.040 | 78.00 | kHz | 4 | 82.00 kHz |
99.00 kHz | 69.960 | 70.040 | 78.00 | kHz | 5 | 82.00 kHz |
10.00 kHz | 69.996 | 70.004 | 7.80 | kHz | 6 | 8.20 kHz |
|
|
|
|
|
|
|
1.00 kHz | 69.9996 | 70.0004 | 0.78 | kHz | 7 | 0.82 kHz |
300.00 Hz1 | 69.99988 | 70.00012 | 225.00 | Hz | 255 .00 Hz | |
1 This is not an EMC analyzer specication; however, the 300 Hz span is tested to 65 % to keep |
the narrow bandwidth accuracy and residual FM measurement uncertainty at | a minimum. If |
the 300 Hz span accuracy is >5% the additional measurement uncertainty may need to be | |
included for the bandwidth accuracy and residual FM measurement uncertainties | . |
Performance Verication Tests