25. Other Input Related Spurious Responses, HP 8593EM
8. Press the following EMC analyzer keys:
4MKR5 More 1 of 3 MARKER ALL OFF
4DISPLAY5 HOLD
4AUTO COUPLE5 AUTO ALL
4SPAN5 1 4MHz5
4AMPLITUDE5 REF LVL 97 4dBV5
ATTEN AUTO MAN (MAN) 0 4dB5
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
Band 1
9.On the EMC analyzer, press 4FREQUENCY5, 4, 4GHz5.
10.Set the synthesized sweeper CW to 4 GHz.
11.Enter the power sensor 4 GHz CAL Factor into the measuring receiver.
12.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
13.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table