58.Tracking Generator Feedthrough, HP 8591EM Option 010
6.Press 4SGL SWP5 , wait for the completion of a new sweep, then press 4MKR !5
MKR ! HIGH .
Subtract the MKR amplitude reading from 87 dBV, then enter the result in the EMC analyzer as the REF LVL OFFSET. For example, if the marker reads 86.79 dBV, enter +0.21 dB.
87 dBV 0 (86.79 dBV) = +0.21 dB
Then press the following EMC analyzer keys:
4AMPLITUDE5 More 1 of 3 More 2 of 3 REF LVL OFFSET (enter
calculated value)
7.Connect one 50 termination to the EMC analyzer INPUT 50 and another to the tracking generator's RF OUT 50 .
8.Press 4AUX/USER5, Track Gen , then SRC PWR ON OFF (OFF).
9.Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 0 4Hz5
4SPAN5 10 4MHz5
4AMPLITUDE5 97 4dBV5
4BW5 AVG BW AUTO MAN (AUTO)
4MKR5 More 1 of 2 MARKER ALL OFF
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
10. Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4MKR !5 More 1 of 3 MARKER !REF LVL
4SPAN5 2 4MHz5
Wait for the AUTO ZOOM message to disappear, then press 4MKR5 More 1 of 3 MK TRACK ON OFF (OFF).
11.Press 4FREQUENCY5 and adjust the center frequency until the LO feedthrough peak is on the
4SPAN5 50 4kHz5
4AMPLITUDE5 57 4dBV5
4BW5 AVG BW AUTO MAN (MAN) 30 4Hz5
Performance Verication Tests