30. Spurious Response, HP 8593EM
15. On the EMC analyzer, press the following keys:
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the peaking message to disappear.
16.Press 4MKR !5 MARKER ! HIGH , More 1 of 3 More 2 of 3 MARKER 1 , then record the measuring receiver reading at 2.8 GHz in Table
17.Set the synthesized sweeper CW to 5.6 GHz.
18.Press the following EMC analyzer keys:
4FREQUENCY5 5.6 4GHz5
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
19.Adjust the synthesized sweeper power level until the Marker 1 Amplitude reads 0 dB 60.20 dB.
20.Enter the power sensor 6 GHz Cal Factor into the measuring receiver.
21.Record the Measuring Receiver Reading at 5.6 GHz in Table
22.Subtract the Measuring Receiver Reading at 5.6 GHz from the Measuring Receiver Reading at 2.8 GHz, then record this value as the Frequency
Response Error (FRE) in Table
at 2.8 GHz is 07.05 dBm, the Frequency Response Error would be 07.05 dBm 0 (06.45 dBm) = 00.60 dB.
Measuring Receiver Reading at 2.8 GHz 0 Measuring Receiver Reading at
5.6GHz = FRE
Performance Verication Tests