30. Spurious Response, HP 8593EM

27.Adjust the synthesized sweeper power level for an EMC analyzer marker amplitude reading of 107 dB￿V 60.2 dB.

28.On the EMC analyzer, press the following keys:

4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)

4MKR !5 MARKER ! HIGH

More 1 of 3 More 2 of 3 MARKER 1

4FREQUENCY5 5.6 4GHz5

4SPAN5 10 4MHz5

29.Remove the ￿lter and connect the synthesized sweeper output directly to the EMC analyzer INPUT 50 ￿.

30.On the EMC analyzer, press the following keys:

4MKR !5 MARKER ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the peaking message to disappear.

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 100 4kHz5

31.Reinstall the ￿lter between the synthesized sweeper output and the EMC analyzer INPUT 50 ￿.

32.Set the EMC analyzer by pressing the following keys:

4AMPLITUDE5 REF LVL 67 dB￿V

4BW5 AVG BW AUTO MAN (MAN) 30 4Hz5

More 1 of 2 VID AVG ON OFF (ON) 10 4ENTER5

4SGL SWP5

Wait until AVG 10 is displayed along the left side of the CRT display.

33.Press 4MKR !5 MARKER ! HIGH , then record the Marker Amplitude Reading as TR Entry 3 of the performance veri￿cation test record.

The Marker Amplitude Reading should be more negative than the speci￿cation previously recorded as TR Entry 2 of the performance veri￿cation test record.

2-194 Performance Veri￿cation Tests