Amplitude Speci￿cations

Relative Quasi-P eak Response to a CISPR Pulse (dB)

(All except Option 703)

 

 

 

 

 

Frequency Band

 

 

 

Pulse Repetition

120 kHz EMI B W

9 kHz EMI B W

200 Hz EMI B W

Frequency (Hz)

0.03 to 1 GHz

0.15 to 30 MHz

10 to 150 kHz

1000

+8.0 6 1.0

+4.5 6 1.0

 

100

0 dB (reference)*

0 dB (reference)*

+4.0

 

6 1.0

60

+3.0

 

6 1.0

25

0 dB (reference)*

20

 

06.5 6 1.0

 

09.0 6 1.0

 

10

014.0

6 1.5

010.0

6 1.5

04.0

 

6 1.0

5

07.5

 

6 1.5

2

026.0

6 2.0

020.5

6 2.0

013.0

6 2.0

1

028.5

6 2.0

022.5

6 2.0

017.0

6 2.0

Isolated Pulse

031.5

6 2.0

023.5

6 2.0

019.0

6 2.0

*Reference pulse amplitude accuracy relative to the CW signal is <1.5 dB as speci￿ed in CISPR Pub. 16.

CISPR reference pulse: 0.044 ￿Vs for 30 MHz to 1 GHz, 0.316 ￿Vs for 15 kHz to 30 MHz, and 13.5 ￿Vs 61.5 ￿Vs for 9 kHz to 150 kHz.

Gain Compression y

 

>10 MHz

￿0.5 dB (total power at input mixer* = 97 dB￿V)

* Mixer Power Level (dB￿V) = Input Power (dB￿V) 0 Input Attenuation (dB).

yIf IF BW ￿ 300 Hz, this applies only if signal separation ￿ 4 kHz and signal amplitudes ￿ Reference Level + 10 dB.

Displayed A verage Noise Level

(Input terminated, 0 dB attenuation,

 

30 Hz AVG BW, sample detector)

1 kHz IF BW

 

400 kHz to 1 MHz

￿08 dB￿V

1 MHz to 1.5 GHz

￿08 dB￿V

1.5 GHz to 1.8 GHz

￿06 dB￿V

30 Hz IF BW

 

400 kHz to 1 MHz

￿023 dB￿V

1 MHz to 1.5 GHz

￿023 dB￿V

1.5 GHz to 1.8 GHz

￿021 dB￿V

 

 

HP 8591EM Speci￿cations and Characteristics 4-7