Agilent Technologies E6380-90073 manual CHANnelPOWerFILTer CHANnelPOWerFILTer?, Darm

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 102
Image 102

GPIB Command Dictionary

CANalyzer subsystem

:CHANnel:POWer:FILTer

:CHANnel:POWer:FILTer?

These commands set/query the filter for the channel power measurement.

‘1.23M’ measures the absolute channel power level of the RF signal in a 1.23 MHz bandwidth centered around the RF channel or tune frequency setting. This measurement must be calibrated using the CAN:CHAN:CPOW:CAL command each time the frequency is changed.

‘30kHz’ measures the adjacent channel power (that is the power in the sidebands around the chosen CDMA channel).

Syntax

CAN:CHAN:POW:FILT ‘1.23M’

‘30kHz’

CAN:CHAN:POW:FILT?

Screen/field equivalent

CAN:CHAN:POW:FILT controls the Ch Pwr Filt field of the CDMA ANALYZER screen while measuring channel power.

:DARM

This command works with the :ARM command to control the measurements.

Syntax

CAN:DARM

Screen/field equivalent

CAN:DARM controls the Disarm subfield of the Analyzer field on the CDMA ANALYZER screen.

102

Chapter 1

Page 102
Image 102
Agilent Technologies E6380-90073 manual CHANnelPOWerFILTer CHANnelPOWerFILTer?, Darm