GPIB Command Dictionary

AFANalyzer subsystem

:GTIMe <real number> :GTIMe?

This command sets the gate time (how long the AF counter samples the input signal) for the audio frequency counter.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

AFAN:GTIM <real number> !valid from 10 milliseconds to 1 second

AFAN:GTIM?

Screen/field equivalent

AFAN:GTIM controls the AF Cnt Gate field on the AF ANALYZER screen.

:INPut <string> :INPut?

These commands set/query the input to the audio analyzer.

Syntax

AFAN:INP ‘FM Demod’

‘AM Demod’

‘SSB Demod’

‘Audio In’

‘Ext Mod’

‘FM Mod’

‘AM Mod’

‘Audio Out’

AFAN:INP?

Screen/field equivalent

AFAN:INP controls the AF Anl In field on the AF ANALYZER screen.