GPIB Command Dictionary
AFG2ENCoder subsystem
:NAMPsNTACs[:FOCC]:RATE <real number> :NAMPsNTACs[:FOCC]:RATE?
These commands set/query the data rate of the encoded message.
This command utilizes the “Real Number Setting Syntax” on page 299 but does not use the :STATe command.
Syntax
AFG2:NAMP:FOCC:RATE
AFG2:NAMP:FOCC:RATE?
Screen/field equivalent
AFG2:NAMP[:FOCC]:RATE controls the Data Rate field on the
SIGNALING ENCODER screen when the Mode field is set to
:NAMPsNTACs[:FOCC]:FILLer:DATA1 <string> :NAMPsNTACs[:FOCC]:FILLer:DATA2 <string> :NAMPsNTACs[:FOCC]:FILLer:DATA1?
:NAMPsNTACs[:FOCC]:FILLer:DATA2?
These commands set/query FOCC filler data. Each filler contains 7 hexadecimal characters representing the 2 type bits and 26 information bits of the control filler/message word. The control filler is sent continuously when the :NAMPsNTACs[:FOCC]:FILL:SEND command is used or after a control message has been sent using :NAMPsNTACs:SEND.
Both filler fields must be full (seven digits) for the forward control channel information to be structured correctly. Do not leave any blank spaces.
To use this command, AFG2:NAMP:CHAN must be set to ‘Cntl’.
Syntax
AFG2:NAMP[:FOCC]:FILL:DATA1 <string>
AFG2:NAMP[:FOCC]:FILL:DATA2 <string>
AFG2:NAMP[:FOCC]:FILL:DATA1?
AFG2:NAMP[:FOCC]:FILL:DATA2?
Screen/field equivalent
AFG2:NAMP[:FOCC]:FILL:DATA1 and DATA 2 control the Filler fields for Stream A and Stream B on the SIGNALING ENCODER screen when the Mode field is set to
Chapter 1 | 93 |