GPIB Command Dictionary

SANalyzer subsystem

:MASK:RELative:UPPer:POINts:EDIT

This command selects/queries the point number to edit on the upper spectrum analyzer mask. The maximum number is 15.

Syntax

SAN:MASK:REL:UPP:POIN:EDIT <integer 1-15>

SAN:MASK:REL:UPP:POIN:EDIT?

Screen/field equivalent

SAN:MASK:REL:UPP:POIN:EDIT controls the EditPt field on the Mask menu of the SPEC ANL screen when the Edit Mask field is set to Upper, and the Mask Type field is set to Rel.

:MASK:RELative:UPPer:POINts:LEVEL1 (through LEVEL15) :MASK:RELative:UPPer:POINts:LEVEL1? (through LEVEL15)

These commands set/query the level of the defined points on the upper spectrum analyzer mask.

Syntax

To define the levels in a four-point mask:

SAN:MASK:REL:UPP:POIN:LEVEL1 <real number>

SAN:MASK:REL:UPP:POIN:LEVEL2 <real number>

SAN MASK:REL:UPP:POIN:LEVEL3 <real number>

SAN:MASK:REL:UPP:POIN:LEVEL4 <real number>

SAN:MASK:REL:UPP:POIN:LEVEL1?

Screen/field equivalent

SAN:MASK:REL:UPP:POIN:LEVEL1 controls the Level Ofs field on the Mask menu of the SPEC ANL screen when the EditPt field is set to 1, the Edit Mask field is set to Upper, and the Mask Type field is set to Rel.

328

Chapter 1

Page 328
Image 328
Agilent Technologies E6380-90073 manual MASKRELativeUPPerPOINtsEDIT