GPIB Command Dictionary

CANalyzer subsystem

:SAMPle:TIME <real number> :SAMPle:TIME?

These commands set/query the amount of time that rho or EVM is measured. This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

CAN:SAMP:TIME <real number>

!values from 0.00025 to 0.02 seconds, for Forward Analyzer Direction (CAN:DIR ‘Fwd’)

!values from 0.0025 to 0.01 seconds for Reverse Channel Direction (CAN:DIR ‘Rev’)

CAN:SAMP:TIME?

Screen/field equivalent

CAN:SAMP:TIME sets the Meas Intvl field of the CDMA ANALYZER screen when a rho or EVM measurement is selected.

:SPECial <string> :SPECial?

These commands set/query the phase of the I and Q signals that are analyzed by the Test Set.

Syntax

CAN:SPEC ‘Normal’ !set for IS-95 standard rotation

‘Inverted’

CAN:SPEC?

Screen/field equivalent

CAN:SPEC controls the setting of the Anl Special field of the CDMA ANALYZER screen.