GPIB Command Dictionary

SANalyzer subsystem

:CFRequency <real number> :CFRequency?

These commands set/query the center frequency of the spectrum analyzer’s display.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

SAN:CFR <real number>

SAN:CFR?

Screen/field equivalent

SAN:CFR controls the Center Freq field on the SPEC ANL screen’s Main menu.

:CONTrol <string> :CONTrol?

These commands set/query the Controls menu for the spectrum analyzer. You can choose controls for the main, RF generator, marker, trigger, mask, or auxilliary functions.

Syntax

SAN:CONT ‘Main’

‘RF Gen’

‘Marker’

‘Auxiliary’

‘Mask’

SAN:CONT?

Screen/field equivalent

SAN:CONT controls the Controls field on the SPEC ANL screen.

314

Chapter 1

Page 314
Image 314
Agilent Technologies E6380-90073 manual CFRequency real number CFRequency?, CONTrol string CONTrol?