GPIB Command Dictionary

MEASure subsystem

:DEC:DTMF:HIGH:FREQuency:ABSolute <meas cmnd> :DEC:DTMF:HIGH:FREQuency:ABSolute?

These command set/query the frequency of the high tone in the DTMF pair. The MEAS:DEC:DTMF:HIGH:FREQ:DISP ‘Freq’ command must be used to display the absolute frequency of the high tone.

Syntax

MEAS:DEC:DTMF:HIGH:FREQ:ABS

!See “Multiple Real Number Setting Syntax” on page 282

MEAS:DEC:DTMF:HIGH:FREQ:ABS? !returns up to 20 real values

Screen/field equivalent

MEAS:DEC:DTMF:HIGH:FREQ:ABS? reads the Hi Tone column on the SIGNALING DECODER screen. The Mode field must be set to DTMF and the upper subfield of the Hi Tone field must be set to Freq.

:DEC:DTMF:HIGH:FREQuency:ERRor <meas cmnd> :DEC:DTMF:HIGH:FREQuency:ERRor?

These command set/query the frequency error of the high tone in the DTMF pair. The MEAS:DEC:DTMF:HIGH:FREQ:DISP ‘Frq Err’ command must be used to display the frequency error of the high tone.

Syntax

MEAS:DEC:DTMF:HIGH:FREQ:ERR

!See “Multiple Real Number Setting Syntax” on page 282

MEAS:DEC:DTMF:HIGH:FREQ:ERR? !returns up to 20 real values

Screen/field equivalent

MEAS:DEC:DTMF:HIGH:FREQ:ERR? reads the Hi Tone column on the SIGNALING DECODER screen. The Mode field must be set to DTMF and the upper subfield of the Hi Tone field must be set to Frq Err.

Chapter 1

263

Page 263
Image 263
Agilent Technologies E6380-90073 manual