GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer:POWer[:VALue]?
NOTE | The values returned represent the relative power of individual Walsh |
| channels not the supplemental channels of which they may be a part. |
| The power of supplemental channels in a Power & Noise measurement |
| must be read directly from the Marker Lvl value on the Code Domain |
| Analyzer screen. |
| Syntax |
| MEAS:CDAN:IS2000:TRAC:POW:POW[:VAL]? |
| Screen/field equivalent |
| MEAS:CDAN:IS2000:TRAC:POW:POW[:VAL]? displays information |
| (top value of each bar) for the power and noise measurement when the |
| CD pwr unit is set to Rel on the Reference menu of the CODE |
| DOMAIN screen when the CDMA standard is set to |
:DEC:AMPSTACS:NBITs?
These commands set/query the number of bits measured by the signaling decoder in
Syntax
MEAS:DEC:AMPS:NBIT?
Example
MEAS:DEC:AMPS:NBIT? !returns an integer value reporting the number of bits measured by the signaling unit.
For example, if the Signaling Decoder displays a value of 928 in the Num of Bits field of the SIGNAL DECODER screen, the value 928 will be returned by the query.
Screen/field equivalent
MEAS:DEC:AMPS:NBIT? reads the Num of Bits measurement field on the SIGNALING DECODER screen when the Mode field is set to
Chapter 1 | 259 |