![](/images/new-backgrounds/1182399/182399511x1.webp)
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer:COMPlex:Q[:VALue]?
:CDANalyzer:IS2000:TRACe:POWer:COMPlex:IS2000?
NOTE | The values returned represent the absolute power of individual Walsh |
| channels not the supplemental channels of which they may be a part. |
| The power of supplemental channels must be read directly from |
| the Q Pwr Level field on the CODE DOMAIN screen when Complex |
| Power measurement has been selected. |
| Syntax |
| MEAS:CDAN:IS2000:TRAC:POW:COMP:Q[:VAL]? |
| MEAS:CDAN:IS2000:TRAC:POW:COMP:IS2000[:VAL]? |
| Screen/field equivalent |
| MEAS:CDAN:IS2000:TRAC:POW:COMP:Q[:VAL]? and |
| MEAS:CDAN:IS2000:TRAC:POW:COMP:IS2000[:VAL]? displays |
| information (bottom set of bars) for the Complex Power measurement |
| when the CD pwr unit is set to Rel on the Reference menu on the |
| CODE DOMAIN screen when the CDMA standard is set to |
:CDANalyzer:IS2000:TRACe:POWer:NOISe:ACTive?
Syntax
MEAS:CDAN:IS2000:TRAC:POW:NOIS:ACT?
Screen/field equivalent
There are no fields on the CODE DOMAIN screen equivalent to the return value for this query. The return information is only accessible when the CDMA standard is set to
256 | Chapter 1 |