GPIB Command Dictionary

MEASure subsystem

:CDANalyzer:IS2000:TRACe:POWer:COMPlex:I[:VALue]?

:CDANalyzer:IS2000:TRACe:POWer:COMPlex:COMBined[:VALue]?

IS-2000 only. This query returns an array of 128 values that represents the Spreader I Input Power of both IS-95 and IS-2000 channels relative to the selected reference power value.

NOTE

The values returned represent the absolute power of individual Walsh

 

channels not the supplemental channels of which they may be a part.

 

The power of supplemental channels must be read directly from

 

I Pwr Level fields on the CODE DOMAIN screen when Complex

 

Power measurement has been selected.

 

Syntax

 

MEAS:CDAN:IS2000:TRAC:POW:COMP:I[:VAL]?

 

MEAS:CDAN:IS2000:TRAC:POW:COMP:COMB[:VAL]?

 

Screen/field equivalent

 

MEAS:CDAN:IS2000:TRAC:POW:COMP:I[:VAL]? and

 

MEAS:CDAN:IS2000:TRAC:POW:COMP:COMB[:VAL]? returns

 

information (top set of bars) displayed for the Complex Power

 

measurement when CD pwr unit is set to Rel on the Reference

 

menu on the CODE DOMAIN screen when the CDMA standard is set to

 

“IS-2000”.