GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer[:BASic]:ALL?
If the selected location is part of a supplemental channel, all the Walsh channels that comprise the supplemental channel and total power for the supplemental channel will be returned.
Syntax
MEAS:CDAN:IS2000:TRAC:POW[:BAS]:ALL?
Screen/field equivalent
There is no equivalent field for this query on the CODE DOMAIN screen. The information is accessible only when a Power, Fast Power, orPower & Noise measurement is being displayed, CD pwr unit is set to Rel from the Reference image on the CODE DOMAIN screen and the CDMA standard is set to
:CDANalyzer:IS2000:TRACe:POWer[:BASic][:VALue]?
NOTE | The values returned represent the power of individual Walsh channels |
| not the supplemental channels of which they may be a part. The power |
| of supplemental channels must be read directly from the Marker Lvl |
| value on the CODE DOMAIN screen. |
| Syntax |
| MEAS:CDAN:IS2000:TRAC:POW[:BAS][:VAL]? |
| Screen/field equivalent |
| MEAS:CDAN:IS2000:TRAC:POW[:BAS][:VAL]? returns values that |
| match those displayed for Power, Fast Power, and Power & Noise |
| measurements when CD pwr unit is set to Rel from the Reference |
| image on the CODE DOMAIN screen when the CDMA standard is set |
| to |
252 | Chapter 1 |