Agilent Technologies E6380-90073 manual AFRequencyFM meas cmnd AFRequencyFM?

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 215
Image 215

GPIB Command Dictionary

MEASure subsystem

:AFRequency:FM <meas cmnd> :AFRequency:FM?

These commands set/query the FM deviation measurement. This measurement is active when FM Mod or FM Demod is selected by using the AFAN:INP command (see page 59) and SNR is not selected by the MEAS:AFR:SEL command (see page 216).

Syntax

MEAS:AFR:FM <meas cmd>

!See “Number Measurement Syntax” on page 283. MEAS:AFR:FM? !returns real value

Example

MEAS:AFR:FM:AVERage:RESet !resets the number of

averages

MEAS:AFR:FM?

!returns the measured

value

Screen/field equivalent

MEAS:AFR:FM? reads the FM Deviation measurement field on the AF ANALYZER screen when the AF Anl In field is set to FM Mod, or FM Demod. This measurement is also displayed on the RF GENERATOR and RF ANALYZER screens.

:AFRequency:FREQuency <meas cmnd> :AFRequency:FREQuency?

These commands set/query the audio frequency measurement. This measurement is active when AF Freq is selected by using the MEAS:AFR:SEL ‘AF Freq’ command (see page 216).

Syntax

MEAS:AFR:FREQ <meas cmd>

!See “Number Measurement Syntax” on page 283.

MEAS:AFR:FREQ? !returns real value

Example

MEAS:AFR:SEL ‘AF Freq’ !selects

the

measurement

MEAS:AFR:FREQ?

!returns

the

measurement’s value

Screen/field equivalent

MEAS:AFR:FREQ? reads the AF Freq measurement field on the AF ANALYZER screen. This measurement is also displayed on the RF GENERATOR and RF ANALYZER screens.

Chapter 1

215

Page 215
Image 215
Agilent Technologies E6380-90073 AFRequencyFM meas cmnd AFRequencyFM?, AFRequencyFREQuency meas cmnd AFRequencyFREQuency?