GPIB Command Dictionary

CONFigure subsystem

:SPOR11SPORT11SB11SP11:PARity <string> :SPOR11SPORT11SB11SP11:PARity?

These commands set/query the parity of the SERIAL 11 port.

Syntax

CONF:SPOR11:PAR ‘None’ ‘Odd’ ‘Even’ ‘Always 1’ ‘Always 0’

CONF:SPOR11:PAR?

Screen/field equivalent

CONF:SPOR11:PAR controls the Parity field of the I/O CONFIGURE screen.

:SPOR11SPORT11SB11SP11:DATA <string> :SPOR11SPORT11SB11SP11:DATA?

These commands set/query the number of data bits per data word sent over the SERIAL 11 port.

Syntax

CONF:SPOR11:DATA ‘7 bits’

‘8 bits’

CONF:SPOR11:DATA?

Screen/field equivalent

CONF:SPOR11:DATA controls the Data Length field of the I/O

CONFIGURE screen.