GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:PNUMber:OFFSet <integer> :IS2000:PNUMber:OFFSet? :IS2000:PNumber:OFFset <integer> :IS2000:PNumber:OFFset?

IS-2000 only. These commands set/query the contents of the PN offset field in the Code Domain analyzer.

This command utilizes the :INCR command of the “Real Number

Setting Syntax” on page 299.

Syntax

CDAN:IS2000:PNUM:OFFS <integer> !values from 0 to 512 CDAN:IS2000:PNUM:OFFS? !returns the value of the PN Offset CDAN:IS2000:PN:OFFS <integer> !values from 0 to 512 CDAN:IS2000:PN:OFFS? !returns the value of the PN Offset

Screen/field equivalent

CDAN:IS2000:PNUM:OFFS and CDAN:IS2000:PN:OFFS control the PN Offset field on the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”. This field is displayed only when the Find PN field on the PN Setup image is set to Manual.

:IS2000:POWer:CHANnel[:CALibrate]

IS-2000 only. This command initiates the channel power calibration. When this command is received, an internally generated calibration signal is measured using the average power technique. Channel power is also measured and a correction factor is generated. This correction factor is applied to subsequent channel power measurements. Calibrations should be performed whenever a new set of measurements is made and whenever the frequency of the measured signal is changed.

Syntax

CDAN:IS2000:POW:CHAN[:CAL]

Screen/field equivalent

CDAN:IS2000:POW:CHAN[:CAL] controls the Chn Pwr Cal measurement field on the Reference image of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000” and the CD pwr unit field is set to Abs.

140

Chapter 1

Page 140
Image 140