GPIB Command Dictionary

CDANalyzer subsystem

:PUNit <string> :PUNit?

These commands set/query whether power measurements are absolute power measurements or relative power measurements.

‘Abs’ - Absolute code domain power displays the power in each of the 64 Walsh channels, relative to the total power inside a 1.23 MHz bandwidth centered at the tune frequency. The 64 Walsh channels (0 through 63) are represented by a vertical bar on the analyzer’s display.

‘Rel’ - Relative code domain power displays the power in each of the 64 Walsh channels, relative to the pilot’s power. (Pilot power is approximately two-thirds of the total power.) The 64 Walsh codes (0 through 63) are represented by a vertical bar on the analyzer’s display.

Syntax

CDAN:PUN ‘Abs’

‘Rel’

CDAN:PUN?

Screen/field equivalent

CDAN:PUN controls the CD pwr unit field on the CODE DOM screen when the CDMA standard is set to “IS-95 Only”.

:SAMPle:TIME <real number> :SAMPle:TIME?

These commands set/query the time interval for the code domain measurements.

Syntax

CDAN:SAMP:TIME <real number> !values 0.00025 to 0.00125 s

CDAN:SAMP:TIME?

Screen/field equivalent

CDAN:SAMP:TIME controls the Meas Intvl field on the Aux menu of the CODE DOM screen when the CDMA standard is set to “IS-95 Only”.