Agilent Technologies E6380-90073 MASKRELativeLOWerPOINtsNUMBer MASKRELativeLOWerPOINtsNUMBer?

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 325
Image 325

GPIB Command Dictionary

SANalyzer subsystem

:MASK:FIXed:UPPer:POINts:FREQuency1 (through FREQ15) :MASK:FIXed:UPPer:POINts:FREQuency1? (through FREQ15)

These commands set/query the frequency (Hz) of the defined points on the upper spectrum analyzer mask.

The units for the mask are GHz, MHz, kHz, and Hz.

Syntax

To define the frequencies in a four-point mask:

SAN:MASK:FIX:UPP:POIN:FREQ1 <real number>

SAN:MASK:FIX:UPP:POIN:FREQ2 <real number>

SAN MASK:FIX:UPP:POIN:FREQ3 <real number>

SAN:MASK:FIX:UPP:POIN:FREQ4 <real number>

SAN:MASK:FIX:UPP:POIN:FREQ1?

Screen/field equivalent

SAN:MASK:FIX:UPP:POIN:FREQ1 controls the Freq field on the Mask menu of the SPEC ANL when the EditPt field is set to 1, the Edit Mask field is set to Upper, and the Mask Type field is set to Fix.

:MASK:RELative:LOWer:POINts:NUMBer

:MASK:RELative:LOWer:POINts:NUMBer?

This command sets/queries the number of points in the lower spectrum analyzer mask. The maximum number is 15.

Syntax

SAN:MASK:REL:LOW:POIN:NUMB <integer 1-15>

SAN:MASK:REL:LOW:POIN:NUMB?

Screen/field equivalent

SAN:MASK:REL:LOW:POIN:NUMB controls the #Pts field on the Mask menu of the SPEC ANL screen when the Edit Mask field is set to Lower, and the Mask Type field is set to Rel.

Chapter 1

325

Page 325
Image 325
Agilent Technologies E6380-90073 manual MASKRELativeLOWerPOINtsNUMBer MASKRELativeLOWerPOINtsNUMBer?