GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:THReshold <real number> :IS2000:THReshold?

IS-2000 Only. These commands set/query the minimum display threshold for code domain measurements. Walsh channels that have power levels below the threshold are not displayed as active channels.

Syntax

CDAN:IS2000:THR <real number> !values minus 80 to 0.0 dB

CDAN:IS2000:THR? !returns the Threshold value

Screen/field equivalent

CDAN:THRS controls the Threshold field on the Aux menu of the

CODE DOM screen when the CDMA standard is set to “IS-2000”.