GPIB Command Dictionary

AFG2ENCoder subsystem

:FILTer <string> :FILTer?

To improve performance, one of four premodulation filters is automatically selected for each encoder mode. The automatically selected filter can be changed only by using GPIB commands.

We recommend that you leave this setting at its automatically selected setting.

If it is necessary to override the automatic settings, the AFG2:FILT:MODE ‘ON’ command must be executed first (filter mode OFF is the power up default state.) The following error will occur if the user attempts to select an alternate filter without first setting the filter mode to ON: Entry not accepted. The syntax to change or query the premodulation filter is shown in the example below.

Syntax

AFG2:FILT ‘None’

‘20 kHzLPF’ ‘250 HzLPF’ ‘150 HzLPF’

AFG2:FILT?

Example

AFG2:FILT:MODE 'ON' !turn filter selection mode on

AFG2:FILT:MODE? !query the current mode setting

AFG2:FILT 'NONE20kHz LPF250Hz LPF150Hz LPF' !select one to change the setting

AFG2:FILT? !query the new filter setting

Screen/field equivalent

AFG2:FILT has no screen/field equivalent.

Chapter 1

69

Page 69
Image 69
Agilent Technologies E6380-90073 manual FILTer string FILTer?, Example