GPIB Command Dictionary
AFG2ENCoder subsystem
:FILTer <string> :FILTer?
To improve performance, one of four premodulation filters is automatically selected for each encoder mode. The automatically selected filter can be changed only by using GPIB commands.
We recommend that you leave this setting at its automatically selected setting.
If it is necessary to override the automatic settings, the AFG2:FILT:MODE ‘ON’ command must be executed first (filter mode OFF is the power up default state.) The following error will occur if the user attempts to select an alternate filter without first setting the filter mode to ON: Entry not accepted. The syntax to change or query the premodulation filter is shown in the example below.
Syntax
AFG2:FILT ‘None’
‘20 kHzLPF’ ‘250 HzLPF’ ‘150 HzLPF’
AFG2:FILT?
Example
AFG2:FILT:MODE 'ON' !turn filter selection mode on
AFG2:FILT:MODE? !query the current mode setting
AFG2:FILT 'NONE20kHz LPF250Hz LPF150Hz LPF' !select one to change the setting
AFG2:FILT? !query the new filter setting
Screen/field equivalent
AFG2:FILT has no screen/field equivalent.
Chapter 1 | 69 |