GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:GAIN:MODE <string> :IS2000:GAIN:MODE?

For IS-2000 only. These commands set/query the mode of the Gain field, allowing you to choose between an autoranging gain (‘Auto’) or a fixed gain (‘Hold’).

This command may conflict with the CDMA analyzer’s settings CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the ‘Auto’ setting. When autoranging is turned off, setting either analyzer’s path gain will automatically change the other’s to the same setting.

For average power measurements, use the CAN:POW:GAIN:MODE command.

Syntax

CDAN:IS2000:GAIN:MODE ‘Auto’

‘Hold’

CDAN:IS2000:GAIN:MODE?

Screen/field equivalent

CDAN:IS2000:GAIN:MODE sets the Auto/Hold subfield of the Gain image of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.

:IS2000:GAIN:PATH <real number> :IS2000:GAIN:PATH?

For IS-2000 only. These commands query/set returns the gain information from the Code Domain analyzer. The command allows you to fine tune the gain setting.

Syntax

CDAN:IS2000:GAIN:PATH <real number> !values from 0 to 36 dB in 0.1 dB steps

CDAN:IS2000:GAIN:PATH?

Screen/field equivalent

CDAN:IS2000:GAIN:PATH has no corresponding field. This applies when the CDMA standard is set to “IS-2000”.

132

Chapter 1

Page 132
Image 132
Agilent Technologies E6380-90073 manual IS2000GAINMODE string IS2000GAINMODE?, IS2000GAINPATH real number IS2000GAINPATH?