GPIB Command Dictionary

CANalyzer subsystem

:ACP:POWer:OFFSet <real number> :ACP:POWer:OFFSet?

This command sets the frequency offset used when making adjacent channel power measurements. The setting designates the frequency offset from the center frequency at which the measurement is made. Valid values are 100 kHz to 3 MHz.

Syntax

CAN:ACP:POW:OFFS <real number>

CAN:ACP:POW:OFFS?

Example

CAN:ACP:POW:OFFS 100 kHz !sets the freq. offset to 100 kHz

CAM:ACP:POW:OFFS?

!returns freq. offset setting

Screen/field equivalent

CAN:ACP:POW:OFFS controls the ACP Offset field on the CDMA ANALYZER screen.

:ARM

This command arms the measurements based on the type of measurement (single or continuous), qualifying events, and triggering events.

To put the CDMA analyzer in single triggering mode, use the

TRIG:MODE:RETR SING command.

For continuous triggering, use TRIG:MODE:RETR REP.

Syntax

CAN:ARM!arms the CDMA analyzer measurements

Example

TRIG:MODE:RETR SING

!sets

the

trigger mode to Single

CAN:ARM

!arms

the

measurement

Screen/field equivalent

CAN:ARM controls the Arm Meas subfield of the Analyzer field.