GPIB Command Dictionary

CGENerator subsystem

:EVENsec <string> :EVENsec?

These commands set/query the even-second input’s setting. When enabled, signals applied to the EVEN SECOND SYNC IN port synchronize the Test Set’s timing circuits.

Syntax

CGEN:EVEN ‘Enable’

‘Not’

CGEN:EVEN?

Screen/field equivalent

CGEN:EVEN controls the Even Sec In field of the CDMA

GENERATOR screen.

:SPECial <string> :SPECial?

These commands set/query the rotation of IQ modulated signals. Normal creates and analyzes IQ signals with standard rotation. Inverted creates and analyzes IQ signals with reverse-rotation.

(Reverse rotation is opposite to the direction specified in the IS-95 standards.)

Syntax

CGEN:SPEC ‘Normal’

‘Invert’

CGEN:SPEC?

Screen/field equivalent

CGEN:SPEC controls the Gen Special field of the CDMA

GENERATOR screen. This field is also displayed on the CDMA

ANALYZER screen