GPIB Command Dictionary

MEASure subsystem

:CDANalyzer:IS2000:TRACe:POWer[:BASic]:ACTive?

IS-2000 only. This query returns the active channel information for the Code Domain trace when making a Power, Fast Power or Fast Power Synchronize measurement. The return entity consists of an integer representing the number of active channels followed by a four value record for each channel. The four values for each record are channel number, spread factor for this channel, Power for this channel and a fourth value that should be ignored.

Syntax

MEAS:CDAN:IS2000:TRAC:POW[:BAS]:ACT?

Screen/field equivalent

There is no equivalent field for this query on the CODE DOMAIN screen. The information is accessible only when a Power, Fast Power, orPower & Noise measurement is being displayed, CD pwr unit is set to Rel from the Reference image on the CODE DOMAIN screen and the CDMA standard is set to “IS-2000”.

Chapter 1

251

Page 251
Image 251
Agilent Technologies E6380-90073 manual CDANalyzerIS2000TRACePOWerBASicACTive?