Agilent Technologies E6380-90073 manual OMODe string OMODe?, OPERationAUTO OPERationHOLD

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 156
Image 156

GPIB Command Dictionary

CONFigure subsystem

:OMODe <string> :OMODe?

These commands set/query the state of the RF offset defined in the

CONF:OFR command.

Syntax

CONF:OMOD ‘On’

‘Off’

CONF:OMOD?

Screen/field equivalent

CONF:OMOD controls the RF Offset field of the INSTRUMENT

CONFIGURE screen.

:OPERation:AUTO

:OPERation:HOLD

These commands control the autoranging routine in the Test Set. :AUTO enables the routines, :HOLD disables them.

Syntax

CONF:OPER:AUTO !enables and autoranging

CONF:OPER:HOLD !disables and autoranging

Screen/field equivalent

CONF:OPER:AUTO and :OPER:HOLD control the Range Hold field of the INSTRUMENT CONFIGURE screen.

:PCMCia:CARD:STATus?

This command queries whether or not a PC card is inserted in the Test Set’s front-panel card slot (No Card, or Inserted).

Syntax

CONF:PCMC:CARD:STAT?

Screen/Field Equivalent

CONF:PCMC:CARD:STAT? queries the Card Status field on the I/O CONFIGURE screen.

156

Chapter 1

Page 156
Image 156
Agilent Technologies E6380-90073 manual OMODe string OMODe?, OPERationAUTO OPERationHOLD, PCMCiaCARDSTATus?