GPIB Command Dictionary

MEASure subsystem

:SANalyzer:MARKer[:NORMal]:FREQuency <meas cmnd> :SANalyzer:MARKer[:NORMal]:FREQuency?

These commands set/query the frequency at the marker on the spectrum analyzer’s trace.

This measurement utilizes the Number Measurement Syntax, but does not use the :METer command.

Syntax

MEAS:SAN:MARK:NORM:FREQ <meas cmd>

!See “Number Measurement Syntax” on page 283.

MEAS:SAN:MARK:NORM:FREQ? !returns real value

Example

MEAS:SAN:MARK:NORM:FREQ? !returns the value

Screen/field equivalent

MEAS:SAN:MARK[:NORM]:FREQ? reads the Freq measurement field on the SPEC ANL screen.

:SANalyzer:MARKer[:NORMal]:LEVel <meas cmnd> :SANalyzer:MARKer[:NORMal]:LEVel?

These commands set/query the RF level at the marker on the spectrum analyzer’s trace.

This measurement utilizes the Number Measurement Syntax, but does not use the :METer command.

Syntax

MEAS:SAN:MARK:NORM:LEV <meas cmd>

!See “Number Measurement Syntax” on page 283.

MEAS:SAN:MARK:NORM:LEV? !returns real value

Example

MEAS:SAN:MARK:NORM:LEV? !returns the value

Screen/field equivalent

MEAS:SAN:MARK[:NORM]:LEV? reads the Lvl measurement field on the SPEC ANL.