GPIB Command Dictionary

MEASure subsystem

:CDANalyzer:TRACe:TOFFset?

IS-95 Only mode. This command queries the timing in the 64 trace points of the code domain analyzer. The measurement returns 64 values (in seconds), one for each point.

Syntax

MEAS:CDAN:TRAC:TOFF?

Screen/field equivalent

MEAS:CDAN:TRAC:TOFF? does not correspond to any field on the CODE DOM screen. This applies when the CDMA standard is set to “IS-95 Only”.

:CDANalyzer:TRACe:PERRor?

IS-95 Only mode. This command queries the phase in the 64 trace points in the code domain analyzer. The measurement returns 64 values (in radians), one for each point.

Syntax

MEAS:CDAN:TRAC:PERR?

Screen/field equivalent

MEAS:CDAN:TRAC:PERR? does not correspond to any field on the CODE DOM screen. This applies when the CDMA standard is set to “IS-95 Only”.

:CDANalyzer:TRACe:TPOWer?

IS-95 Only mode. This command queries the absolute (true) power in each of the 64 Walsh channels, relative to the total power inside a 1.23 MHz bandwidth centered at the tune frequency. The measurement returns 64 values (in dB), one for each point.

Syntax

MEAS:CDAN:TRAC:TPOW?

Screen/field equivalent

MEAS:CDAN:TRAC:TPOW? does not correspond to any field on the CODE DOM screen. This applies when the CDMA standard is set to “IS-95 Only”.

Chapter 1

233

Page 233
Image 233
Agilent Technologies E6380-90073 manual CDANalyzerTRACeTOFFset?, CDANalyzerTRACePERRor?, CDANalyzerTRACeTPOWer?