GPIB Command Dictionary

AFG2ENCoder subsystem

:NAMPsNTACs[:FOCC]:FM <real number> :NAMPsNTACs[:FOCC]:FM?

This command sets/queries the data level when the AFG2:DEST ‘FM’ command is used. The data level units are kHz.

This command utilizes the “Real Number Setting Syntax” on page 299.

Syntax

AFG2:NAMP:FOCC:FM

AFG2:NAMP:FOCC:FM?

Screen/field equivalent

AFG2:NAMP[:FOCC]:FM controls the Data Level field when AFGen2 To field is set to FM on the SIGNALING ENCODER screen when the Mode field is set to NAMP-NTAC, and the Channel field is set to Cntl.

:NAMPsNTACs[:FOCC]:LEVel <real number> :NAMPsNTACs[:FOCC]:LEVel?

These commands set/query the data level when the AFG2:DEST ‘Audio Out’ command is used. The data level units are mV.

This command utilizes the “Real Number Setting Syntax” on page 299.

Syntax

AFG2:NAMP:FOCC:LEV

AFG2:NAMP:FOCC:LEV?

Screen/field equivalent

AFG2:NAMP[:FOCC]:LEV controls the Data Level field when AFGen2 To field is set to Audio Out on the SIGNALING ENCODER screen when the Mode field is set to NAMP-NTAC, and the Channel field is set to Cntl.