GPIB Command Dictionary
MEASure subsystem
:SANalyzer:MARKer:DELTa:FREQuency <meas cmnd> :SANalyzer:MARKer:DELTa:FREQuency?
These commands set/query the delta marker frequency measurement. This measurement is the frequency of the delta marker minus the frequency of the normal marker. If the delta marker is to the right of the normal marker, the delta frequency is positive. If the delta marker is to the left of the normal marker, the delta frequency is negative.
The units for this measurement are GHz, MHz, kHz, and Hz.
This measurement utilizes the Number Measurement Syntax, but does not use the :METer command.
Syntax
MEAS:SAN:MARK:DELT:FREQ <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:SAN:MARK:DELT:FREQ? !returns real value
Example
SAN:MARK:PEAK
MEAS:SAN:MARK:DELT:FREQ? !returns the value
Screen/field equivalent
MEAS:SAN:MARK:DELT:FREQ? reads the Delta Mrkr, Freq measurement field on the SPEC ANL screen when Norm/Delta field is set to Delta on the Marker controls menu.