CAUTION

Always use the three-prong ac power cord supplied with this product.

 

Failure to ensure adequate earth grounding by not using this cord may

 

cause personal injury and/or product damage.

 

This product is designed for use in Installation Category II and

 

Pollution Degree 2 per IEC 1010 and IEC 664 respectively. For indoor

 

use only.

 

This product has autoranging line voltage input, be sure the supply

 

voltage is within the specified range.

 

Ventilation Requirements: When installing the product in a cabinet, the

 

convection into and out of the product must not be restricted. The

 

ambient temperature (outside the cabinet) must be less than the

 

maximum operating temperature of the product by 4° C for every 100

 

watts dissipated in the cabinet. If the total power dissipated in the

 

cabinet is greater than 800 watts, then forced convection must be used.

 

To prevent electrical shock, disconnect instrument from mains (line)

 

before cleaning. Use a dry cloth or one slightly dampened with water to

 

clean the external case parts. Do not attempt to clean internally.

Product

CE - the CE mark is a registered trademark of the European Union. A

Markings

CE mark accompanied by a year indicated the year the design was

 

proven.

 

CSA - the CSA mark is a registered trademark of the Canadian

 

Standards Association.