Agilent Technologies E6380-90073 manual TRACeNORMalize string TRACeNORMalize?, TRACeSAVE

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 335
Image 335

GPIB Command Dictionary

SANalyzer subsystem

:TRACe:NORMalize <string> :TRACe:NORMalize?

These commands set/query the type of trace display. ‘A Only’ provides a continuously updated display (normal operation). ‘A-B’ displays the difference between the trace saved using SAN:TRAC:SAVE and the currently displayed trace.

Syntax

SAN:TRAC:NORM ‘A Only’

‘A-B’

SAN:TRAC:NORM?

Screen/field equivalent

SAN:TRAC:NORM controls the A Only/A-Bsubfield of the Normalize field on the tracking generator of the SPEC ANL screen’s Auxilliary menu.

:TRACe:SAVE

This command stores the trace that is currently displayed on the spectrum analyzer display.

Syntax

SAN:TRAC:SAVE

Screen/field equivalent

SAN:TRAC:SAVE controls the Save B subfield of the Normalize field on the SPEC ANL screen’s Auxilliary menu.

Chapter 1

335

Page 335
Image 335
Agilent Technologies E6380-90073 manual TRACeNORMalize string TRACeNORMalize?, TRACeSAVE