GPIB Command Dictionary

SANalyzer subsystem

:SPAN <real number> :SPAN?

These commands set/query the span of the spectrum analyzer display.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

SAN:SPAN <real number>

SAN:SPAN?

Screen/field equivalent

SAN:SPAN controls the Span field on the Main menu SPEC ANL screen.

:TGENerator:AMPLitude <real number> :TGENerator:AMPLitude?

These commands set/query the amplitude of the tracking generator. To select the tracking generator use the SAN:RFG ‘Track’ command.

This command utilizes the “Real Number Setting Syntax” on page 299.

Syntax

SAN:TGEN:AMPL <real>

SAN:TGEN:AMPL?

Screen/field equivalent

SAN:TGEN:AMPL controls the Amplitude field on the RF Gen menu of the SPEC ANL screen. The Track/Fixed field must be set to Track.

Chapter 1

331

Page 331
Image 331
Agilent Technologies E6380-90073 manual Span real number SPAN?, TGENeratorAMPLitude real number TGENeratorAMPLitude?