GPIB Command Dictionary

SANalyzer subsystem

:MARKer[:NORMal]:NPEak

This command moves the marker to the next peak of the spectrum analyzer’s measurement trace.

Syntax

SAN:MARK:NORM:NPE

Screen/field equivalent

SAN:MARK:NORM:NPE controls the Marker To, Next Peak field on the Marker menu of the SPEC ANL screen when Norm/Delta field is set to Norm on the Marker controls menu.

:MARKer[:NORMal]:NPLevel <real number> :MARKer[:NORMal]:NPLevel?

These commands set/query the level of the next peak signal.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

SAN:MARK:NORM:NPL <real number>

SAN:MARK:NORM:NPL?

Screen/field equivalent

SAN:MARK:NORM:NPL controls the Level field on the SPEC ANL screen when Norm/Delta field is set to Norm on the Marker controls menu.

:MARKer[:NORMal]:PEAK

This command moves the marker to the highest peak of the signals on the spectrum analyzer display trace.

Syntax

SAN:MARK:NORM:PEAK

Screen/field equivalent

SAN:MARK:NORM:PEAK controls the Marker To, Peak field on the Marker menu of the SPEC ANL screen when Norm/Delta field is set to Norm on the Marker controls menu.