GPIB Command Dictionary

AFG2ENCoder subsystem

:AMPSTACS:SAT:FM <real number> :AMPSTACS:SAT:FM?

These commands set/query the SAT level when the AFG2:DEST ‘FM’ and AFG2:AMPS:CHAN ‘Voice’ commands are used. The SAT level’s units are kHz.

This command utilizes the “Real Number Setting Syntax” on page 299.

Syntax

AFG2:AMPS:SAT:FM

AFG2:AMPS:SAT:FM?

Screen/field equivalent

AFG2:AMPS:SAT:FM controls the SAT Level field when AFGen2 To field is set to FM on the SIGNALING ENCODER screen when the Mode field is set to AMPS-TACS and the Channel field is set to Voice.

:AMPSTACS:SAT:LEVel <real number> :AMPSTACS:SAT:LEVel?

These commands set/query the SAT level when the AFG2:DEST ‘Audio Out’ and AFG2:AMPS:CHAN ‘Voice’ commands are used. The SAT level units are mV.

This command utilizes the “Real Number Setting Syntax” on page 299.

Syntax

AFG2:AMPS:SAT:LEV

AFG2:AMPS:SAT:LEV?

Screen/field equivalent

AFG2:AMPS:SAT:LEV controls the SAT Level field when AFGen2 To field is set to Audio Out on the SIGNALING ENCODER screen when the Mode field is set to AMPS-TACSand the Channel field is set to Voice.

82

Chapter 1

Page 82
Image 82
Agilent Technologies E6380-90073 Ampstacssatfm real number AMPSTACSSATFM?, AMPSTACSSATLEVel real number AMPSTACSSATLEVel?