GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:TRIGger:ARM

IS-2000 only. This command arms code domain measurements. This command does not affect continuous measurement and will always be overridden by GPIB triggering commands. To effectively put the code domain analyzer in single triggering mode, use the TRIG:MODE:RETR SING command. For continuous triggering, use TRIG:MODE:RETR REP. See “TRIGger subsystem” on page 346 for a more complete discussion of GPIB triggering.

Syntax

CDAN:IS2000:TRIG:ARM !arms the measurement

Screen/field equivalent

CDAN:IS2000:TRIG:ARM controls the Arm Meas subfield of the Analyzer field on the Main menu of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.

:IS2000:TRIGger:DARM

IS-2000 only. This command disarms the code domain measurements. This command does not affect continuous measurement and will always be overridden by GPIB triggering commands. To effectively put the code domain analyzer in single triggering mode, use the TRIG:MODE:RETR SING command. For continuous triggering, use TRIG:MODE:RETR REP. See “TRIGger subsystem” on page 346 for a more complete discussion of GPIB triggering.

Syntax

CDAN:IS2000:TRIG:DARM !disarms measurements that have been previously armed

Screen/field equivalent

CDAN:IS2000:TRIG:DARM controls the Disarm subfield of the Analyzer field on the Main menu of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.