GPIB Command Dictionary

AFANalyzer subsystem

:NOTCh:GAIN <string> :NOTCh:GAIN?

These commands set/query the gain of the AF analyzer’s notch filter amplifier. This command is only effective when autoranging is turned off using the AFAN:RANG ‘Hold’ command.

Gain is automatically set when the AFAN:RANG ‘Auto’ command is used. If autoranging is enabled (AFAN:RANG ‘Auto’) and you attempt to change the gain setting by sending the AFAN:NOTC:GAIN command, the value you send will be overridden by the autoranging function.

Syntax

AFAN:NOTC:GAIN ‘0 dB’

‘10 dB’

‘20 dB’

‘30 dB’

‘40 dB’

AFAN:NOTC:GAIN?

Screen/field equivalent

AFAN:NOTC:GAIN controls the Notch Gain field on the AF

ANALYZER screen when the Gain Cntl field is set to Hold.

:NOTCh:FREQuency <real number> :NOTCh:FREQuency?

These commands set/query the center frequency for the variable frequency notch filter. This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

AFAN:NOTC:FREQ <real number> !valid 330 Hz to 10 kHz

AFAN:NOTC:FREQ?

Screen/field equivalent

AFAN:NOTC:FREQ controls the Notch Freq field on the AF

ANALYZER screen.