GPIB Command Dictionary

SANalyzer subsystem

:MARKer[:NORMal]:POSition <real number> :MARKer[:NORMal]:POSition?

These commands set/query the position of the marker on the spectrum analyzer.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

SAN:MARK:NORM:POS <real number>

SAN:MARK:NORM:POS?

Screen/field equivalent

SAN:MARK:NORM:POS controls the Position field on the Marker menu of the SPEC ANL screen when Norm/Delta field is set to Norm on the Marker controls menu.

:MARKer:REFmode

:MARKer:REFmode?

These commands set/query the spectrum analyzer’s reference marker mode when using the delta markers.

When the Ref Mrkr field is set to Hold, the reference marker is frozen at its current position (frequency and level) regardless of changes in the signal level or center frequency.

When you switch from Hold to Norm, the reference marker will stay at its current horizontal setting (frequency), but will track the level of the incoming signal.

Syntax

SAN:MARK:REF ‘Norm’

‘Hold’

SAN:MARK:REF?

Screen/field equivalent

SAN:MARK:REF controls the Ref Mrkr field on the Marker menu of the SPEC ANL.

320

Chapter 1

Page 320
Image 320