GPIB Command Dictionary

RFGenerator subsystem

:FREQuency <real number> :FREQuency?

These commands set/query the frequency of the RF generator. The Test Set must first be in frequency tuning mode. (See CONF:RFD ‘Freq’ on page 164).

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

RFG:FREQ <real number>

RFG:FREQ?

Screen/field equivalent

RFG:FREQ controls the RF Gen Freq field on the RF GENERATOR screen, when in frequency tuning mode. This field is also displayed on the spectrum analyzer’s RF Gen menu.

:MODulation:AOUT <string> :MODulation:AOUT?

These commands set/query the coupling between the demodulated audio and the AUDIO OUT port.

Syntax

RFG:MOD:AOUT ‘AC’

‘DC’

RFG:MOD:AOUT?

Screen/field equivalent

RFG:MOD:AOUT controls the Audio Out field on the RF GENERATOR screen.