GPIB Command Dictionary

RFANalyzer subsystem

:IFBW <string> :IFBW?

These commands set/query the IF filter bandwidth.

Syntax

RFAN:IFBW ‘15 kHz’

‘230 kHz’

RFAN:IFBW?

Screen/field equivalent

RFAN:IFBW controls the IF Filter field on the RF ANALYZER screen.

:INPut <string> :INPut?

These commands set/query the RF input port.

Note: Some measurements can only be made on one port.

Syntax

RFAN:INP ‘RF In’

‘Ant’

RFAN:INP?

Screen/field equivalent

RFAN:INP controls the RF Input Port field on the RF ANALYZER. This field is also displayed on the SPEC ANL, CODE DOM, and CDMA ANALYZER screens.

:PMEasurement:DETector <string> :PMEasurement:DETector?

These commands set/query the analog TX power measurement method.

Syntax

RFAN:PME:DET ‘Peak’

‘Sample’

RFAN:PME:DET?

Screen/field equivalent

RFAN:PME:DET controls the TX Pwr Meas field on the RF ANALYZER screen.