GPIB Command Dictionary

AFG2ENCoder subsystem

:NAMPSNTACS[:FOCC]:MESSage:DATA1 <string> :NAMPSNTACS[:FOCC]:MESSage:DATA2 <string> :NAMPSNTACS[:FOCC]:MESSage:DATA1?

:NAMPSNTACS[:FOCC]:MESSage:DATA2?

These commands set/query FOCC Stream A and Stream B message data. Like the filler data, the message information can only be input in full (seven digit) lines. Also, message streams A and B must have the same number of lines in them.

To use this command, AFG2:NAMP:CHAN must be set to ‘Cntl’.

Syntax

AFG2:NAMP[:FOCC]:MESS:DATA1 <string>

AFG2:NAMP[:FOCC]:MESS:DATA2 <string>

AFG2:NAMP[:FOCC]:MESS:DATA1?

AFG2:NAMP[:FOCC]:MESS:DATA2?

Screen/field equivalent

AFG2:NAMP:MESS controls the Message fields for FOCC Stream A and Stream B on the SIGNALING ENCODER screen when the Mode field is set to NAMP-NTAC. These fields are displayed only when the Channel field is set to Cntl.

:NAMPsNTACs:FVC:MESS <string> :NAMPsNTACs:FVC:MESS?

These commands set/query the seven hexadecimal characters (28 bits) of the FVC message. These 7 characters are combined with 12 parity bits calculated by the encoder to output a 40-character data stream. This data is output when the AFG2:NAMP:FVC:SEND ‘Message’ command is sent.

The entire field must contain data. No blank spaces are allowed. The DSAT is turned off while the FVC message stream is sent.

Syntax

AFG2:NAMP:FVC:MESS <string>

AFG2:NAMP:FVC:MESS?

Screen/field equivalent

AFG2:NAMP:FVC:MESS controls the FVC Message field on the

SIGNALING ENCODER screen when the Mode field is set to

NAMP-NTAC, and the Channel field is set to Voice.