Agilent Technologies E6380-90073 IS2000GAINVALue?, IS2000MARKerMODE string IS2000MARKerMODE?

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 134
Image 134

GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:GAIN:VALue?

IS-2000 only. This query returns the gain value, as a string, that is shown in the gain field of the code domain screen. This command will only be enabled when autoranging is turned on (CDAN:IS2000:GAIN:MODE ‘Auto’).

Syntax

CDAN:IS2000:GAIN:VAL? !returns 0 dB, 6 dB, 12 dB, 18 dB, 24 dB, 30 db, 36 dB

Screen/field equivalent

CDAN:IS2000:GAIN:VAL? reads the lower subfield of the Gain image of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000” and the upper field is set to Auto.

:IS2000:MARKer:MODE <string> :IS2000:MARKer:MODE?

IS-2000 only. These commands set/query the string that matches the field contents of the field that selects the way channel width information is presented: In terms of Channel size or Walsh order.

Syntax

CDAN:IS2000:MARK:MODE <string>

CDAN:IS2000:MARK:MODE?

Screen/field equivalent

CDAN:IS2000:MARK:MODE? reads the selection field Chan Size/Walsh Ord/Sprd Fact/Rate RC-3/RateRC-4/RateRC-5in the measurement portion of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.

134