Agilent Technologies E6380-90073 manual CDANalyzerMARKerTIME meas cmnd CDANalyzerMARKerTIME?

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 231
Image 231

GPIB Command Dictionary

MEASure subsystem

:CDANalyzer:MARKer:TIME <meas cmnd> :CDANalyzer:MARKer:TIME?

IS-95 Only mode. These commands set/query the time at the marker on the code domain analyze’s trace during a timing measurement (see CDAN:MEAS on page 124). This measurement is made relative to the value at Walsh Code 0.

This measurement utilizes the Number Measurement Syntax, but does not use the :METer command.

Syntax

MEAS:CDAN:MARK:TIME <meas cmd>

!See “Number Measurement Syntax” on page 283. MEAS:CDAN:MARK:TIME?

Example

MEAS:CDAN:MARK:TIME?

Screen/field equivalent

MEAS:CDAN:MARK:TIME? reads the Time field on the Marker menu of the CODE DOM screen when a timing measurement is selected. This applies when the CDMA standard is set to “IS-95 Only”.

:CDANalyzer:PNOFfset?

IS-95 Only mode. This command queries the PN offset measurement.

Syntax

MEAS:CDAN:PNOF?

Example

MEAS:CDAN:PNOF?

Screen/field equivalent

MEAS:CDAN:PNOF? queries the PN Ofs measurement field on the CODE DOM screen whent he CDMA standard is set to “IS-95 Only”, and the CD pwr unit is set to Abs on the CD Setup image of the CODE DOMAIN screen.