GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:MEASurement:MODE <string> :IS2000:MEASurement:MODE?

IS-2000 only. These commands set/query the measurement state for the code domain analyzer. This mode can be overridden with the TRIG:MODE:RETR REP and TRIG:MODE:RETR SING trigger commands. When the trigger commands are used, the state of the field on the display are not updated. For example, the screen may show Single selected when TRIG:MODE:RETR REP is used.

Syntax

CDAN:IS2000:MEAS:MODE ‘Single’

‘Cont’

CDAN:IS2000:MEAS:MODE?

Screen/field equivalent

CDAN:IS2000:MEAS:MODE controls the Single/Cont subfield of the Measurement field of the Main image of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.

:IS2000:MEASurement:ORDer <string> :IS2000:MEASurement:ORDer?

IS-2000 only. These commands set/query the channel order of measurement currently shown on the CODE DOMAIN screen. The channel order can follow the Hadamard ordering in which all 128 channels are ordered consecutively from 0 to 127, or they can follow the Bit Reverse ordering.

Syntax

CDAN:IS2000:MEAS:ORD ‘Bit Reverse’

‘Hadamard’

CDAN:IS2000:MEAS:ORD?

Screen/field equivalent

CDAN:IS2000:MEAS:ORD controls the Meas Order field of the Aux image on the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.