GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:PNUMber:MODE <string> :IS2000:PNUMber:MODE? :IS2000:PNumber:MODE <string> :IS2000:PNumber:MODE?

IS-2000 only. These commands set/query the PN setup.

‘Auto’ is used to calculate the system’s PN offset if it is unknown. When ‘Auto’ is selected, the CDAN:IS2000:PNUM:INCR command can be used to enter a PN increment and speed the PN offset calculation.

‘Manual’ is used to enter a known PN offset. When ‘Manual’ is selected, the PN offset is entered using the CDAN:IS2000:PN:OFFS command.

Syntax

CDAN:IS2000:PNUM:MODE ‘Auto’

‘Manual’

CDAN:IS2000:PNUM:MODE? !returns the PN Offset mode

CDAN:IS2000:PN:MODE ‘Auto’

‘Manual’

CDAN:IS2000:PN:MODE? !returns the PN Offset mode

Screen/field equivalent

CDAN:IS2000:PNUM:MODE and CDAN:IS2000:PN:MODE control the Find PN field on the PN Setup image of the CODE DOMAIN screen when the CDMA standard is set to “IS-2000”.